version 2.0 rev 18 Feb 2021
68
contact Andor Product Support to arrange a pre-emptive Zyla chamber service.
There is some anticipated variability as to when moisture ingress can be expected to present as an issue to
camera performance following saturation of the desiccant, based on variability in environmental humidity and
airowaroundthecamera.Forexample,whileacameramaybeoperatedinadryair-conditionedlaboratory
environmentthroughouttheworkingday,thelaboratoryairconditioningmaybeturnedoovernight,during
which the camera chamber can become exposed to high heat and humidity. Under such circumstances, back-
lledcamerascanbemoreexposedtoelevatedratesofmoistureingressintothesensorchamberandpre-
emptive chamber service may be considered on a more frequent basis.
5.5 Cooling Hoses and Connections
The user should routinely check all coolant hoses and connections for signs of leakage, damage or wear. All
seals must be intact before powering on camera system and any worn / damaged items must be replaced
immediately.