RELATED FUNCTIONS
74 / 101 NTA47230-300A6
TC-TM 66XX
Remarks:
The parameters and scaling points are saved when turning the unit off.
Beware, when changing functions (thermocouple or resistive sensor) or units, scaling goes OFF. The data defined in the
point lists are then erroneous.
D.2 Differential measurements
The relative measurement function available on the device makes it possible to cancel a constant or spurious value via
programming.
When the relative measurement function is enabled, the symbol is displayed on the measurement screen.
The NULL/define menu makes it possible to program the value of the Tare (positive or negative). This value is obtained
from the measurements:
Value Displayed = Value measured Value of the Tare