EasyManua.ls Logo

AOIP TC6621 - Data Acquisition and Storage; Acquisition and Storage Procedures

AOIP TC6621
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
RELATED FUNCTIONS
78 / 101 NTA47230-300A6
TC-TM 66XX
D.5 Storage of acquisitions in progress
The TC6621 is designed to store 10,000 values in one or more acquisition bursts.
Using the F2 key, enter the Mode menu.
Select the Memory function.
Confirm using the VAL key.
The drop-down list displays the following functions:
REC.MEASUREMENT:
Enable the triggering of an acquisition on a case-by-case basis.
If an acquisition has already been opened, then the following screen is displayed:
If you wish to save it, press the F2 key (YES) to confirm.
You are then requested to enter the name of a file. Using the navigation keys ( and ), scroll down the letters.
Using the navigation keys ( and ), move the cursor by one position.
Using the F2 key( ), you can delete the characters entered
Once you have entered the file name, confirm by pressing the VAL key.
RUN:
Launches the storage of data following the parameters set in the “parameters” function. The pictogram is displayed on the
measurement screen
STOP:
Stops the storage in progress.

Table of Contents

Related product manuals