5 Conductivity Measurement
Press key to turn on the tester. Rinse the probe in distilled water and dry it.
Dip the probe in sample solution, stir gently, and allow it to stand still in the solution until a stable
reading is reached. Get readings after comes up and stays. Press to switch from Conductivity
to TDS, and Salinity
Notes
a) The TDS and Salinity measurements are converted from the conductivity
measurements via a certain conversion factor.
b) The tester can calibrate 84μS, 1413 μS and 12.88 mS conductivity calibration
solution. User can conduct 1 to 3 points calibration. Refer to the table below. Usually
calibrating the tester with 1413 μS conductivity buffer solution alone shall meet the
testing requirement.
Calibration Indication Icon
Calibration Standards
84 μS/cm
1413 μS/cm
12.88 mS/cm
5.1 The tester has been calibrated before leaving the factory. Generally, users can use the tester directly
or users can test conductivity buffer solutions first. If the error is large, then calibration is needed.
5.2 For conductivity calibration solutions, we recommend that users replace new solutions after 5 to 10
times of use to keep the standard solution’s accuracy. Do NOT pour the used calibration solutions back
into the solution bottles in case of contamination.
5.3 Temperature compensation factor: The default setting of the temp. compensation factor is 2.0%/.
User can adjust the factor based on test solution and experimental data in parameter setting P4.
Solution
Temperature
compensation factor
Solution
Temperature
compensation factor
NaCl
2.12%/˚C
10% Hydrochloric acid
1.32%/˚C
5% NaOH
1.72%/˚C
5% Sulfuric acid
0.96%/˚C
Dilute ammonia
1.88%/˚C
5.4 TDS and conductivity is linear related, and its conversion factor is 0.40-1.00. Adjust the
factor in parameter setting P5 based on the requirements in different industries. The factory

Other manuals for Apera Instruments EC60

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Apera Instruments EC60 and is the answer not in the manual?

Apera Instruments EC60 Specifications

General IconGeneral
BrandApera Instruments
ModelEC60
CategoryTest Equipment
LanguageEnglish

Related product manuals