25
Appendix
Manual for Doctor mode (for physician or licensed practitioner use only)
It is possible for a physician or licensed practitioner to see the compliance meter, to reset the compli-
ance meter or to lock/unlock the treatment parameters. This is achieved using doctor mode.
1. Enter to Doctor mode: Press and hold on ‘TIME’ key, then turn on the unit with channel 1/2
knob. If you can see the ‘hr’ symbol show on the LCD (the rate is 0 Hz), you are in the doctor
mode. (Some model with ‘min’ symbol)
2. See Compliance meter value: After you enter the doctor mode, the compliance
meter value is showing on the LCD with ‘hr’ symbol. (Some models with ‘min’ symbol)
3. Reset the compliance meter value: When the doctor want to reset the compliance
meter’s value for a new treatment cycle, you can do as the follow steps:
When you are in the doctor mode, press ‘MODE’ key first, then press the reset key. Thus, the
value will be reset.
*resetkey:‘Rate▼’forTENS/EMS/Combo/IF/HV;‘▼’forEMS(C).