EasyManua.ls Logo

APPA 605 - Electrical Specifications

APPA 605
18 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
14
Electrical Specifications
Accuracy is ±(% reading + number of digits) at 23°C ± 5°C
< 80%RH.
Voltage Measurement
Function Range Accuracy
DCV 600.0V
±(1%+5dgt)
ACV 600.0V
±(1.5%+5dgt)(50~60Hz)
±(2%+5dgt)(61~500Hz)
LPF ACV 600.0V
±(1.5%+5dgt)(50~60Hz)
±(5%+5dgt)(61~400Hz)
Start measuring voltage : AC 0.6V.
Over voltage protection : 600V rms or dc.
The cut-off frequency of the low pass filter : 1 kHz.
Input Impedance : 3M // less than 100pF.
CMRR / NMRR : (Common Mode Rejection Ratio)
(Normal Mode Rejection Ratio)
V
AC
: CMRR > 60dB at DC, 50Hz / 60Hz
V
DC
: CMRR > 100dB at DC, 50Hz / 60Hz
NMRR > 50dB at DC, 50Hz / 60Hz
AC Conversion Type :
AC conversions are ac-coupled, true rms responding,
calibrated to the sine wave input.
For non-sine wave add the following Crest Factor
corrections:
For Crest Factor of 1.4 to 2.0, add 1.0% to accuracy.
For Crest Factor of 2.0 to 2.5, add 2.5% to accuracy.
For Crest Factor of 2.5 to 3.0, add 4.0% to accuracy.
CF 3 @ 330V
2 @ 500V
Dimensions (W x H x D) :
95mm x207mm x 52mm with holster.
Accessories : Battery (installed), Test leads and user manual.
Power Requirements : 1.5V x 4 IEC LR6 or AA size.
Pollution degree : 2
EMC : EN 61326-1
Shock vibration : Sinusoidal vibration per MIL-T- 28800E
(5 ~ 55 Hz, 3g maximum).
Drop Protection :
4 feet drop to hardwood on concrete floor.
Indoor Use.