Quick Start Guide
AquaLab Pre Water Activity Meter
Question and Answer IconNeed help?

Do you have a question about the Aqua Lab PRE and is the answer not in the manual?

Aqua Lab PRE Specifications

General IconGeneral
BrandAqua Lab
ModelPRE
CategoryMeasuring Instruments
LanguageEnglish

Summary

The Setup

Instrument Placement and Power On

Place the instrument on a level surface in a stable environment and connect the power cord.

Instrument Warm-up Procedure

Allow the instrument to warm up for 15 minutes before proceeding with testing.

Daily Verification - Primary Standard

Primary Standard Selection and Preparation

Choose a standard close to the sample's activity and prepare it in a sample cup.

Initiating Primary Standard Measurement

Place the prepared sample into the instrument drawer and start the measurement.

Primary Standard Reading and Verification

Verifying Primary Standard Accuracy

Take two readings; they should be within +/-0.01 of the standard's specified value.

Daily Verification - Determine Second Standard

Performing Two-Point Verification

Prepare and measure the secondary standard to confirm instrument accuracy.

Sample Test Preparation

Measurement Results and Best Practices

Avoid Instrument Movement During Measurement

Do not tip or move the instrument with a sample inside to prevent sensor contamination.

Maintenance

Annual Calibration Service

Annual service includes calibration, firmware update, cleaning, warranty, and new supplies.

Troubleshooting Common Error Messages

Mirror Contamination Warning

Warning indicating potential contamination of the mirror sensor.

AquaLink 4 Data Management Software