Default settings ULUF 60/120/400/400-D/500/500-D/700/700-D/800/800-D:
Description
Arctiko default
settings
Customer
Custom Settings (Password)
0000
Set point
Freezer
-40,0
Alarm Settings:
Alarm delay
Freezer
15 min
Door open alarm
Freezer
Enabled
High temp. alarm
Freezer
-30
Low temp. alarm
Freezer
-50
Probe/eprom failure
Freezer
Enabled
Power failure
Freezer
Enabled
Logging time interval
Freezer
1
Description
Arctiko default
settings
Customer
settings
Advanced Settings (Password)
0000
Calibration
Freezer
0,0
Automatic defrost
Freezer
0
Manual defrost
Freezer
OK
Hysteresis
Freezer
2,0
Temp. range limits
Freezer Max.
-20,0
Freezer Min.
-40,0
42
Default settings ULUF 15/65/125:
Description
Arctiko default
settings
Customer
settings
Custom Settings (Password)
0000
Set point
Freezer
-80,0
Alarm Settings:
Alarm delay
Freezer
15 min
Door open alarm
Freezer
Enabled
High temp. alarm
Freezer
-70
Low temp. alarm
Freezer
-90
Probe/eprom failure
Freezer
Enabled
Power failure
Freezer
Enabled
Logging time interval
Freezer
1
Description
Arctiko default
settings
Customer
settings
Advanced Settings (Password)
0000
Calibration
Freezer
0,0
Automatic defrost
Freezer
0
Manual defrost
Freezer
OK
Hysteresis
Freezer
2,0
Temp. range limits
Freezer Max.
-40,0
Freezer Min.
-85,0