52
Electrical Data: EU and North America
Voltage 230 V (EU)
120 V (North America)
Frequency 50 Hz (EU)
60 Hz (North America)
Power consumption (maximum) 900 VA
Number of phases Single phase AC
Fuses 230 V 10 A
120 V 13 A
Protective earth Yes
Potential equalisation Yes
Power supply cord replaced by qualified personnel
Mains disconnection device
Permanently installed in wall, visible and accessible at all times
Mains transient voltage 4 kV*
Creepage distance 3 mm*
Air clearance 1.8 mm*
Direction of movement Up and down, right and left,
clockwise and
counterclockwise
* If local regulations are higher, they must be met.
WARNING
To avoid electric shock, make sure that the equipment is connected to:
• Continuously powered supply mains with protective earth.
• Separate fuse and ground fault circuit interrupter (GFCI)
• A mains disconnection device
• Equipotential bonding point
All installations must be in accordance with local codes and regulations.