System 2000
04.AR.12_15EN • 07/2019
INSTRUCTIONS FOR USE

Questions and Answers

  • D
    Dr. Tammy RobersonAug 13, 2025
    Why is the water flow slow from the shower and filling spout in Arjo Bathtubs?
    • P
      Priscilla NobleAug 13, 2025
      If the water flow from both the shower and filling spout of your Arjo Bathtubs is slow, check the incoming hot water line to see if it is hot. Also, inspect and clean the filter for the incoming water according to the Care and Preventive Maintenance section on page 52.
  • C
    chenstevenAug 17, 2025
    How to fix slow water flow to the shower in Arjo Bathtubs?
    • S
      Sandra HughesAug 17, 2025
      If the water flow to the shower in your Arjo Bathtubs is slow, inspect and clean the shower head. Refer to the Care and Preventive Maintenance section on page 52 for instructions.
  • J
    James Rush DDSAug 21, 2025
    What to do if there's water on the ground near my Arjo Bathtubs?
    • M
      morganronaldAug 21, 2025
      If you find water on the ground near your Arjo Bathtubs, immediately check for leaks in the incoming water lines and the bath drainage hose. Additionally, the bath may have short-circuited, causing it to fill continuously. In either case, turn off both the electrical and water supplies to the unit.
  • M
    Michael DiazAug 25, 2025
    Why won't the auto fill function stop filling my Arjo Bathtubs?
    • S
      Sean Reed IIAug 26, 2025
      If the auto-fill function on your Arjo Bathtubs doesn't stop, it might be due to residue buildup on the mechanical sensor plugs. Refer to the Cleaning and Disinfection Instructions on page 40 for guidance.
  • L
    Lucas CarterAug 28, 2025
    What to do if the fill or shower function stops on my Arjo Bathtubs?
    • H
      hsuttonAug 28, 2025
      If the fill or shower function on your Arjo Bathtubs stops working, reset the scalding protection. Follow the instructions for 'Reset scalding protection (P220 only) (5 steps) on page 18' or 'Reset scalding protection (P300 only) (4 steps) on page 19'.
  • M
    Matthew BriggsSep 1, 2025
    What to do if disinfection does not work on Arjo Bathtubs?
    • A
      Amanda DavidsonSep 1, 2025
      If the disinfection process isn't working on your Arjo Bathtubs, ensure the bathtub is completely empty. Also, make sure that shower, fill or Hydromassage system, Hydrosound system are not activated. If the LED on the control panel is flashing and beeping P300 only, it means the disinfection level in the bottle is low and needs replacement. See section Cleaning and Disinfection Instructions on page 40!
  • J
    josecastroSep 5, 2025
    Why is the sound not working on my Arjo Bathtubs Sound & Vision system, but the light is?
    • D
      Dennis BellSep 5, 2025
      If the sound isn't working on your Arjo Bathtubs' Sound & Vision system, but the light is, the USB memory stick might be defective. If possible check the USB LED light for indication: Solid light power 21 Flashing light reading! Try using a different USB memory stick.
  • B
    bellkellieSep 8, 2025
    What to do if the Sound & Vision system hand control is not working on Arjo Bathtubs?
    • J
      jimmyrobertsSep 8, 2025
      If the hand control for the Sound & Vision system in your Arjo Bathtubs isn't working, check if the LED on the remote control flashes when active. If it doesn't, replace the Sound & Vision system hand control.

Summary

Safety Instructions

General Safety Precautions

Critical safety precautions to avoid injury, fire, shock, and falls.

Preparation Steps

Parts Designation

Control Panels

P220 Control Panel Overview

Description and function of the P220 control panel buttons and display.

P300 Control Panel Overview

Description and function of the P300 control panel buttons and display.

Buttons and Displays

Product Functions: Drain and Autofill

Product Functions: Temperature and Scalding Protection

Scalding Protection System

How the scalding protection works and safety warnings.

Product Functions: Height Adjustment and Disinfection Unit

Hydromassage System Operation

Using the Hydromassage System

Step-by-step guide to operating the Hydromassage system.

Hydrosound System Overview

Using the Hydrosound System

Step-by-step guide to operating the Hydrosound system.

Hydrosound System Controls

Sound & Vision System

Changing Default Settings (P300)

Default Setting Adjustment Procedure

Steps to change default settings for various bath functions.

Bathing with Lift Hygiene Chair: Preparation

Preparations for Hygiene Chair Bathing

Initial steps for preparing the bath with a lift hygiene chair.

Bathing with Lift Hygiene Chair: Transfer

Patient Transfer Steps

Step-by-step instructions for transferring a patient using a lift hygiene chair.

Bathing with Lift Hygiene Chair: Process

Patient Showering: Preparation and Handle Use

Patient Showering: Temperature and Stopping

Ending the Bath and Disinfection

Disinfecting After Each Patient Use

Instructions for disinfecting the bath and hygiene chair after each use.

Bathing with Lift Bath Trolley: Preparation

Preparations for Trolley Bathing

Initial steps for preparing the bath with a lift bath trolley.

Water Temperature Safety Checks

How to check and adjust water temperature safely.

Bathing with Lift Bath Trolley: Transfer

Patient Transfer with Trolley

Steps for transferring a patient using a lift bath trolley.

Bathing with Lift Bath Trolley: Process

Bathing the Patient

The core step of bathing the patient.

Patient Showering with Trolley: Preparation

Post-Bathing and Shower Setup

Steps after bathing and preparing for showering.

Shower Handle and Water Flow Control

How to use the shower handle and control water flow.

Water Temperature and Shower Safety

Safely checking water temperature and showering the patient.

Patient Showering with Trolley: Process and Ending

Water Temperature Verification

How to check and adjust water temperature for patient comfort.

Disinfection After Patient Use

Disinfection Procedure

Instructions for disinfecting the bath and lift bath trolley after each use.

General Cleaning and Disinfection Instructions

Disinfection Safety Warnings

Safety warnings for disinfection to prevent cross-contamination and irritation.

Disinfectant Bottle Replacement

Disinfectant Bottle Replacement Steps

Step-by-step guide to replacing the disinfectant bottle.

Disinfection with Spray Handle: Cleaning

Removing Visual Residue

Steps to remove visible residue from the bath and accessories.

Cleaning Bath Components

Detailed steps for cleaning the bath and accessories.

Disinfection with Spray Handle: Surfaces

Spraying Surfaces and Overflow Outlet

How to spray disinfectant on internal surfaces and the overflow outlet.

Rinsing and Wiping Procedures

Steps for rinsing the bath and wiping contact areas.

Disinfection with Spray Handle: Final Steps

Disinfecting Drains and Final Rinsing

How to disinfect drains and perform final rinsing and drying.

Disinfection with Spray Bottle: Cleaning

Removing Residue with Spray Bottle

Steps to remove visible residue using a spray bottle.

Cleaning Bath Components with Spray Bottle

Detailed steps for cleaning using a spray bottle.

Disinfection with Spray Bottle: Surfaces

Spraying Surfaces with Spray Bottle

How to spray disinfectant using a spray bottle on surfaces and overflow.

Rinsing and Wiping with Spray Bottle

Steps for rinsing and wiping contact areas using a spray bottle.

Disinfection with Spray Bottle: Final Steps

Disinfecting Drains and Final Rinsing (Spray Bottle)

How to disinfect drains and perform final rinsing and drying.

Hydromassage System Disinfection (Automatic)

Hydromassage System Disinfection (Manual)

Hydromassage System Disinfection: Final Steps

Care and Preventive Maintenance Schedule

Caregiver Maintenance Obligations

Schedule for caregiver obligations and maintenance checks.

Equipment Modification Warning

Warning against modifying the equipment.

Caregiver Maintenance Tasks

Monthly Maintenance and Filter Cleaning

Qualified Personnel Maintenance

Annual Service and Calibration

Requirements for annual servicing, calibration, and technician contact.

Qualified Personnel Maintenance Warning

Warning that maintenance must be done by qualified personnel.

Troubleshooting Common Issues

Water Flow and Disinfection Problems

Troubleshooting steps for water flow and disinfection issues.

Leakage and Electrical Faults

Troubleshooting for water leakage and electrical faults.

P300 Error Code Diagnostics

P300 Error Codes and Causes

Explanation of error codes, their causes, and remarks for P300.

Technical Specifications: Site and Electrical

Arjo System 2000 Series Specifications

General IconGeneral
BrandArjo
ModelSystem 2000 Series
CategoryBathtubs
LanguageEnglish

Related product manuals