Sara Plus
KKX52180M-EN_21 • 0/2020
INSTRUCTIONS FOR USE

Other manuals for Arjo SARA PLUS

Questions and Answers

  • H
    huffmarisaAug 19, 2025
    What to do if Arjo Medical Equipment Sara Plus hoist moves unexpectedly?
    • C
      Casey WalkerAug 19, 2025
      If the hoist moves unexpectedly, and releasing the buttons does not work, push the red Emergency stop button and remove the battery. Use the System failure lowering override to put the patient back into a safe seated position, then remove the Sara Plus from use. Do not use the equipment again until it has been inspected and passed for safe working by an Arjo approved service engineer.
  • K
    kedwardsAug 20, 2025
    How to correctly position a patient in an Arjo Sling?
    • A
      Ashley HughesAug 21, 2025
      To correctly position a patient in the Arjo Sling, ensure the clip/lift cords are attached at the right length, and the straps aren't twisted or knotted. Position the patient inside the sling, using the label as a reference, and ensure they are centered. Check that there are no creases on the leg flaps and that they are attached at the correct length. Verify the clip isn't caught. Use the sling according to the allowed combinations. Adjust the patient’s position using the clips/lift cords, or inner clips, and adjust the support belt for comfort. Ensure the patient is holding onto the handles on the lift.
  • J
    Joshua WalkerAug 23, 2025
    What to do if patient feels discomfort in leg area with Arjo Sling?
    • A
      Anthony PerryAug 23, 2025
      If a patient feels discomfort in the leg area when seated in the Arjo Sling, make sure the clip/lift cords are attached at the right length. Also, check that the straps are not twisted or that there are no knots on the lift cords. Ensure the patient is centered in the sling, and that there are no creases on the leg flaps. Finally, make sure the leg flaps are attached at the right length.
  • C
    Cassandra BoydAug 26, 2025
    What to do if Arjo Sling slides up under patient's arms?
    • J
      Jeffrey BrownAug 26, 2025
      If the Arjo Sling is sliding up under the patient’s arms, ensure the patient is positioned on the inside of the sling, using the label for reference. Also, make sure the patient is centered in the sling. Check that there are no creases on the leg flaps and ensure the leg flaps are attached at the right length.
  • N
    Nicole GoldenAug 27, 2025
    How to handle a soiled Arjo disposable sling?
    • M
      Michele HudsonAug 27, 2025
      If the disposable Arjo sling is soiled, stained, or wet, throw it away and replace it with a new one.
  • A
    Angel ManningAug 30, 2025
    What to do if 'Do NOT wash' label is damaged on Arjo Sling?
    • M
      Michael DoyleAug 30, 2025
      If the “Do NOT wash” label is damaged or missing on the disposable Arjo sling, and the underlying red label “Do NOT use/lift” is visible, discard the sling and replace it with a new one.
  • D
    David JohnstonSep 1, 2025
    What to do if Arjo Flites Sling is soiled, stained, or wet?
    • D
      daniel23Sep 1, 2025
      If the Arjo Flites Sling becomes soiled, stained, or wet, dispose of it and replace it with a new one.
  • T
    Trevor WrightSep 3, 2025
    What to do if Arjo Medical Equipment Sara Plus support arms do not move upwards and overload is displayed?
    • L
      Lonnie HoltSep 3, 2025
      If, when pressing the “Raise” button, the Sara Plus makes a noise, “overload” is displayed, and the buzzer beeps continuously but the resident support arms do not move upwards, it may be that the resident support arms are blocked by an obstruction. Remove the obstruction and check the Sara Plus thoroughly for damage before continuing the lifting cycle. If in doubt, use the System failure lower override to return the resident to a safe seated position, then remove the Sara Plus from use. Do not use the equipment again until it has been inspected and passed for safe working by an Arjo approved service engineer.
  • S
    Steven MoyerSep 6, 2025
    What to do if Arjo Medical Equipment Sara Plus chassis legs do not open and overload is displayed?
    • S
      sharon89Sep 6, 2025
      If, when the “Chassis legs open” button is pressed, the Sara Plus makes a noise, “overload” is displayed, and the buzzer beeps continuously but the chassis legs do not open, it may be that the chassis legs are blocked by an obstruction. Remove the obstruction and check the Sara Plus thoroughly for damage before continuing the lifting cycle. If in doubt, use the System failure lower override to return the resident to a safe seated position, then remove the Sara Plus from use. Do not use the equipment again until it has been inspected and passed for safe working by an Arjo approved service engineer.
  • T
    Timothy TaylorSep 8, 2025
    What to do if Arjo Medical Equipment Sara Plus does not raise or lower with the hand control?
    • A
      Antonio MillerSep 8, 2025
      If the Sara Plus does not raise or lower and the chassis legs cannot be opened or closed when using the hand control, the hand control may be damaged. Try operating the Sara Plus with the Dual up/down control located on the mast. If the equipment functions correctly when using these controls, the hand control should be replaced.

Summary

Intended Use and Patient Assessment

Intended Use of Sara Plus

Defines the primary functions and applications of the Sara Plus patient lift.

Safety Instructions and Warnings

General Operational Safety

Emphasizes caregiver training, component checks, and adherence to instructions for safe operation.

Hazard Awareness and Weight Limits

Highlights potential hazards like strangulation and specifies maximum lifting capacities.

Product Description: Controls and Features

Power, Emergency Stop, and Override Functions

Details operation of power buttons, emergency stop, and system failure lower override.

Product Description: Key Components and Adjustments

Product Description: Accessories and Slings

Compatible Slings for Sara Plus

Lists the approved sling types for use with the Sara Plus lift.

How to Use the Sara Plus: General Operation

How to Use the Sara Plus: Standing Sling Transfer

Attaching the Standing Sling

Detailed procedure for connecting the standing sling to the lift using cords.

How to Use the Sara Plus: Completing Standing Sling Transfer

How to Use the Sara Plus: Transfer;Walking Sling Transfer

Attaching the Transfer;Walking Sling

Steps for connecting the transfer/walking sling to the lift.

How to Use the Sara Plus: Completing Transfer;Walking Sling Transfer

How to Use the Sara Plus: Transfer;Walking Sling Walking

Sling Attachment for Walking Practice

Steps for attaching the sling for walking practice and initial adjustments.

How to Use the Sara Plus: Concluding Walking Practice

Battery Instructions: Charging and Indicators

Battery Charging Safety Precautions

Crucial warnings and guidelines for safe battery charging practices.

Battery Discharge Indicator Function

Explains the visual indicator for battery charge level and its behavior.

Battery Instructions: Charging Procedure

Battery Pack Charging Steps

Detailed, step-by-step instructions for charging the Sara Plus battery pack.

Cleaning and Disinfection: General Care

Routine Cleaning Procedures

Guidelines for cleaning the lift using a damp cloth and mild detergent.

Approved Disinfectants and Methods

Lists compatible disinfectants and essential steps for effective disinfection.

Care and Preventive Maintenance: Schedule

Preventive Maintenance Checklist

A comprehensive schedule of checks and actions for caregivers to maintain the lift.

Care and Preventive Maintenance: Checks and Servicing

Daily, Weekly, and General Checks

Essential checks covering before use, daily, weekly, and general lift condition.

Troubleshooting Common Issues

Lift Not Functioning or Slow Operation

Addresses problems like the lift not working or operating slowly, including battery checks.

Control and Movement Blockages

Solutions for issues where controls are unresponsive or movements are blocked.

Overload, OverHeat, and Unexpected Movements

Covers error indicators, noises, and responses to unexpected hoist movements.

Technical Specification: Weights and Electricals

Patient Weight Limits and Component Weights

Details safe working loads for slings and weights of lift components.

Arjo SARA PLUS Specifications

General IconGeneral
Safe Working Load200 kg
Battery Type24V
Total Height1200 mm (47.2 ")
Leg HeightAdjustable
Knee Pad Height AdjustmentAdjustable
Lift MechanismElectric
Height AdjustmentElectric
Wheel Configuration4 castors
FootplateAdjustable