What to do if Arjo Medical Equipment Sara Plus hoist moves unexpectedly?
- CCasey WalkerAug 19, 2025
If the hoist moves unexpectedly, and releasing the buttons does not work, push the red Emergency stop button and remove the battery. Use the System failure lowering override to put the patient back into a safe seated position, then remove the Sara Plus from use. Do not use the equipment again until it has been inspected and passed for safe working by an Arjo approved service engineer.