58
Technical Specifications
$IXOOWHFKQLFDOGHVFULSWLRQRIWKHSystem 2000FDQEHIRXQGLQWKHVHUYLFHGRFXPHQWDWLRQSOHDVHFRQWDFW$UMR
IRUDFRS\&RQWDFWLQIRUPDWLRQDSSHDUVRQWKHODVWSDJHRIWKLV,)8
Site Requirements
NOTE
All drainage, water and electrical installation work should be carried out by qualified personnel, in
accordance with local and national regulations.
Electric supply, Europe and North America
Voltage 230/240 V (Europe)
120 V (North America)
100 V (Japan)
220 V (Korea)
230 V (Saudi Arabia)
Number of phases Single phase AC
Frequency 50 Hz (Europe)
60 Hz (North America)
50/60 Hz (Japan)
60 Hz (Korea)
60 Hz (Saudi Arabia)
Power consumption (maximum) 1100 VA
Fuse 6 A (Europe)
10 A (North America)
12 A (Japan)
6 A (Korea)
6 A (Saudi Arabia)
Protective earth Yes
Potential equalisation Yes
Power supply cord Replaceable by qualified
personnel
WARNING
To avoid electric shock, make sure that the equipment is connected to:
• Continuously powered supply mains with protective earth.
• Separate fuse and ground fault circuit interrupter (GFCI)
• A mains disconnection device
• Equipotential bonding point
All installations must be in accordance with local codes and regulations.