EasyManuals Logo

Arjo System 2000 Series User Manual

Arjo System 2000 Series
72 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #58 background imageLoading...
Page #58 background image
58
Technical Specifications
$IXOOWHFKQLFDOGHVFULSWLRQRIWKHSystem 2000FDQEHIRXQGLQWKHVHUYLFHGRFXPHQWDWLRQSOHDVHFRQWDFW$UMR
IRUDFRS\&RQWDFWLQIRUPDWLRQDSSHDUVRQWKHODVWSDJHRIWKLV,)8
Site Requirements
NOTE
All drainage, water and electrical installation work should be carried out by qualified personnel, in
accordance with local and national regulations.
Electric supply, Europe and North America
Voltage 230/240 V (Europe)
120 V (North America)
100 V (Japan)
220 V (Korea)
230 V (Saudi Arabia)
Number of phases Single phase AC
Frequency 50 Hz (Europe)
60 Hz (North America)
50/60 Hz (Japan)
60 Hz (Korea)
60 Hz (Saudi Arabia)
Power consumption (maximum) 1100 VA
Fuse 6 A (Europe)
10 A (North America)
12 A (Japan)
6 A (Korea)
6 A (Saudi Arabia)
Protective earth Yes
Potential equalisation Yes
Power supply cord Replaceable by qualified
personnel
WARNING
To avoid electric shock, make sure that the equipment is connected to:
• Continuously powered supply mains with protective earth.
• Separate fuse and ground fault circuit interrupter (GFCI)
• A mains disconnection device
• Equipotential bonding point
All installations must be in accordance with local codes and regulations.