19
Technical specification
Component Weights
kg lbs
Tenor complete with scale, battery .............................................70 155
Tenor complete without scale .....................................................69.1 152.3
Maximum total weight of lift (lift + patient) ...................................390 860
In normal use, the Tenor can be broken up into two parts:
Battery pack.................................................................................3.8 8.4
Tenor (without Battery pack)........................................................65.3 144
Tenor with scale (without Battery pack) .......................................66.4 146.4
The Tenor is not intended to be broken up into more than these parts.
Maximum Lifting Capacities (Safe Working Load - SWL)
Tenor ............................................................................... 320 705
All slings - Please verify with the SWL on the sling label
Electrical
Battery type.................................................................................. Rechargeable - sealed lead acid
Battery part number ..................................................................... NDA0100-20
Battery capacity ........................................................................... 4Ah
Lift - Protection class - Ingress of fluid......................................... IPX4
Lift - Protection class - Electrical shock protection....................... Internally powered equipment
Lift nominal voltage...................................................................... 24V
Fuse - Overload ........................................................................... 16A (thermal cutout)
Fuse - PCBA................................................................................ 20A (time delay)
Operating forces - Handset.......................................................... 2.5N
See Battery Charger IFU 001-24257 for technical specifications
Medical Equipment – type protection against electrical shock in accordance with IEC 60601-1.
Certified by TUV for standard IEC 60601-1:2012(ed.3.1) and EN ISO10535:2006.
Duty cycle Max volts Max amps
Lift actuator .................................................................. 10% – 1 min / 9 min 24 V 10 A
“V” Chassis actuator .................................................... 10% – 2 min / 18 min 24 V 2 x 2,3 A
Mode of operation: intermittent
Environment
Air humidity .................................................................. 80% @ 20°C (68°F)
Usage temperature range (ambient)............................ + 5°C (41°F) to + 35°C (95°F)
Optimum usage temperature (ambient)....................... + 20°C (68°F) to + 25°C (77°F)
Storage and transportation temp (ambient) ................. - 10° C (14°F) to + 45°C (113°F)
End of Life Disposal
• All batteries in the product must be recycled separately. Batteries are to be disposed in accordance
with national or local regulations.
• Slings including stiffeners/stabilizers, padding material, any other textiles or polymers or plastic materi-
als etc. should be sorted as combustible waste.
• Lift systems having electrical and electronic components or an electrical cord should be disassembled
and recycled per Waste of Electrical and Electronic Equipment (WEEE) or in accordance with local or
national regulation.
• Components that are primarily be made up of different kinds of metal (containing more than 90% metal
by weight) for example sling bars, rails, upright supports, etc., should be recycled as metals.