What to do if Arkray AUTION ELEVEN AE-4022 Measuring Instruments show 'Elapse of QC deadline'?
- AanthonygarrettSep 5, 2025
If the QC deadline has elapsed, perform control measurements.
What to do if Arkray AUTION ELEVEN AE-4022 Measuring Instruments show 'Elapse of QC deadline'?
If the QC deadline has elapsed, perform control measurements.
What to do if Arkray AUTION ELEVEN AE-4022 show 'Backup memory error'?
If the instrument was unused for a long period, the memory backup battery was exhausted and the measurement results stored in the memory were lost. Keep the instrument powered for at least 11 hours to charge the backup battery. After charging the instrument, set the date and time.
What to do if Arkray Measuring Instruments show 'No paper in the printer'?
If the thermal recording paper has run out, install a new roll of thermal recording paper. If the thermal recording paper roll was not properly installed, make sure the thermal recording paper is installed correctly.
What to do if Arkray Measuring Instruments show 'Auto start sensor error'?
If a test strip was placed while the carrying arm was moving to the suction ports, remove the test strip. If the detection section of the auto start sensor was soiled with urine, clean the detection section.
What to do if Arkray AUTION ELEVEN AE-4022 Measuring Instruments show 'No test strip on the feeder'?
If the feed lever is not attached, turn OFF the instrument and open the maintenance cover. Mount the feed lever. If you find any obstacles or scattered test strips inside, remove them.
What to do if Arkray Measuring Instruments show 'Elapse of operator ID effective time'?
If the operator ID effective time elapsed, enter operator ID.
Sample Type | Urine |
---|---|
Power Supply | AC100-240V, 50/60Hz |
Sample Loading | Automatic |
Display | LCD screen |
Connectivity | USB, RS-232 |
Weight | Approx. 60 kg |
Measurement Method | Reflectance Photometry |
Measurement Items | pH, ORP, Conductivity, Temperature |