Routine Care
Angel
®
cPRP System Operator’s Manual 6-5
Download software from this source” Button
Touching the “Download software from this sourceButton
begins the installation process for the selected software version.
When the button is touched, the selected software version is copied to the Angel System’s
temporary memory and is validated. It is not installed, until the operator confirms the installation
from the Install Screen. Refer to “New Software” on page 6-1 for details.
“Re-check sources” Button
Touching the “Re-check sources” Button will cause the Angel System to
check for any storage device connected to the USB Port for valid software to
install. This is used if the source was not found, or was not properly connected when the
Check for new softwareButton was touched
Visual Inspection
The Angel System should be inspected periodically for any problems such as bent or broken switches,
frayed or twisted power cords, and loose or missing hardware. If the device displays any of the above
conditions, discontinue use of the Angel System until the problem is corrected and it has been verified
that the device is operating correctly.
Verify that the USB Port located on the exterior of the Angel System, is intact and not damaged.
Verify that the Stator Arm is not loose on its pivot pin. It should only rotate up and down with hand
pressure, and have no side-to-side motion. If the arm is loose, lift it to the raised braking position and
locate the cap screw near the base of the Stator Arm. Using a 9/64 inch hex wrench, tighten the cap
screw against the pivot pin.
Routine Care
External Surfaces
Angel System are to be thoroughly cleaned and disinfected according to the following instructions..
Note: In the event of minor fluid (blood) spills that only affect the exterior of the Angel System (and do
not accumulate in the Centrifuge Well), such spills should be addressed within 10 minutes of the
occurrence, if possible, and should not be allowed to dry on the exterior surfaces of the unit. For minor
fluid (blood) spills, ensure that the pump rotor is removed for cleaning and disinfection processes per
the following.
1. Remove the Pump Rotor from the rotor housing by pushing down and turning the rotor counter-
clockwise.