Assemblad Infragas-196 and Infragas-196 PC-Module * Operating manual
ASSEMBLAD S.r.l. - Costruzioni Elettroniche - Via Marzabotto, 4 - 50013 Campi Bisenzio - Firenze - Italy
Tel. +39-055-890485 - Fax +39-055-890496 - E-mail: info@assemblad.it - Web: www.assemblad.it
13
2) Insert take-up tip into the motorcar exhaust pipe, connect RPM probe to one of the
spark plugs cables and insert the temperature probe, if provided, into the oil
measuring rod receptacle.
3) After warm-up, instrument is ready to work; user may select, using FUNC key, one of
the auxiliary parameters available on FUNCTION display.
4) For gas measurement user presses MEAS key causing a self-calibration, after which
analyzer passes to a gas exhaustion stage. During this period the instrument, if gas is
present, periodically performs calibrating procedures (every 10 min.), while
automatically turns the pump off in case of gas lack for at least 10 min.
5) Operator may decide every time to print out data using P1 key, provided that pump is
on.
6) Once measurement has been completed, or whenever the operator wants, he may
decide to terminate gas exhaustion pressing MEAS key. However, if no action is
taken the instrument automatically turns the pump off after approx. 10 min. of
continuous operation, to preserve the pump from an useless stress.
2.4 CALIBRATION USING GAS BOTTLE
Manufacturer recommends a calibration with certified gas bottle (see Periodic
Maintenance Utility chapter) every twelve months. In some countries legislation defines
specific rules. Please contact our distributor for this periodical service activity.
2.5 OXYGEN TRANSDUCER REPLACEMENT
The oxygen transducer (*) life-time is 12 months (minimum) from the date of
delivery. Please contact our distributor for periodical replacement.
If the oxygen measurement, as well as the Lambda factor, are not shown during the
gas analysis it is the time to replace the oxygen sensor.
2.6 NITRIC OXIDE TRANSDUCER REPLACEMENT
The Nitric Oxide transducer (*) life-time is 18 months. It starts from the date of
delivery. Please contact our distributor for periodical replacement.