Do you have a question about the Associated Research HYPOT III 3765 and is the answer not in the manual?
Type | Hipot Tester |
---|---|
Frequency Range | 50/60 Hz |
Output Power | 500 VA |
Display | LCD |
Weight | 15 kg |
AC Hipot Voltage | 0 - 5 kVAC |
DC Hipot Voltage | 6 kV |
AC Hipot Current | 20 mA |
DC Hipot Current | 0 to 10 mA |
Safety Standards | UL, CSA |
Power Supply | 115/230 VAC, 50/60 Hz |
Key safety actions including station survey, personnel control, protocol familiarization, and test awareness.
Crucial safety procedures for operating electrical testers to prevent shock.
Set the correct input line voltage (115 or 230 VAC) on the rear panel.
Connect the NEMA style power cord to the instrument and a grounded source.
Insert the Interlock Disable Key into the signal input connector.
Turn the instrument power switch ON after setup is complete.
Steps to configure instrument parameters: select MENU, then SYSTEM, edit values using soft keys.
Enable Ground Continuity testing by setting the instrument to AC or DC Hipot mode.
Steps to adjust test parameters like Type, Voltage, Limits, Ramp, and Dwell times.
Procedure for connecting products with two or three-prong line cords using an adapter box.
Connect ground return and HV leads to exposed metal chassis of the DUT.
Wrap non-exposed chassis in foil and connect return lead to foil.
Initiate testing by pressing the green TEST button after DUT connection.
Understand Pass (beep) and Fail (alarm) indications displayed on the screen.