English
Ö
The thermodisinfector is equipped with a
product container with relave pump (1 Fig.
15) for the detergent liquid. On request, it
can be equipped with a second container
(2/3 Fig. 15) used for the neutralising liquid
or rinse aid.
a
WARNING. The program can be interrupted
at any me. However, it must be remembered
that once interrupted, it must restart from
the beginning. Only if it is interrupted in the
drying phase can the cycle be considered
completed successfully. The instruments
must be dried.
Ö
IMPORTANT: In the EcoDrying version, the
machine will interrupt the cycle at the end
of the disinfecon. It is the responsibility of
the user to dry the instruments that have
just been washed, using suitable cloths
and following the instrucons contained in
the instrucon manual for each individual
instrument washed. The manufacturer
declines any liability for contaminaon of the
instruments during the manual drying stage.
Ö
IMPORTANT: In the EcoDrying version,
the instruments must NOT be le inside
the machine, at the end of the cycle,
wet, for longer than 5 minutes. This is to
prevent the proliferaon of bacteria and
microorganisms.
Ö
Open the door immediately aer the end of
the program to avoid condensaon forming.
Ö
Hollow instruments with a small internal
diameter must be dried further with air.
Ö
Do not force to open the door to avoid
damaging the device and/or the leak of toxic
steam.
Ö
Parcularly large instruments can be
very hot at the end of the program. Allow