EasyManuals Logo

Athena A228 User Manual

Default Icon
32 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #22 background imageLoading...
Page #22 background image
22
A228/A248/A228I/A248I User’s manual
Electrical fast
transient/burst
IEC 61000-4-4
2 kV for power
supply lines
1 kV for
input/output lines
2 kV for
power
supply
lines
1 kV
for
input/out
put lines
Mains power quality
should be that of a typical
commercial or hospital
environment.
Surge
IEC 61000-4-5
1 kV line(s) to
line(s)
2 kV line(s) to
earth
1 kV
line(s) to
line(s)
2 kV
line(s) to
earth
Mains power quality
should be that of a typical
commercial or hospital
environment.
interruptions and
voltage
variations on
power supply
input lines
IEC
61000-4-11
<5% UT
(>95% dip inUT)
for 0,5 cycle
40% UT
(60% dip in UT)
for 5 cycles
70% UT
(30% dip in UT)
for 25 cycles
<5 %UT
(>95% dip in
<5% UT
(>95%
dip in
UT)
for 0,5
cycle
40% UT
(60% dip
in UT)
for 5
cycles
70% UT
(30% dip
in UT)
for 25
cycles
Mains power quality
should be that of a typical
commercial or hospital
environment.
If the user of the model
A228/A248/A228I/A248I
requires continued
operation during power
mains interruptions, it is
recommended that the
models
A228/A248/A228I/A248I
be powered from an
uninterruptible power
supply or a battery.

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Athena A228 and is the answer not in the manual?

Athena A228 Specifications

General IconGeneral
BrandAthena
ModelA228
CategoryTouch Panel
LanguageEnglish