Listing 15.1: JSON le setting the IP to 172.16.2.4
Please note the UDP port used by the fusionTrack (the 3509 value) cannot currently be changed.
15.2 Changing the local UDP port
The local UDP port (i.e. the random port on the machine used to communicate with the UDP server on
the fusionTrack device) can be forced to a value. To be valid, the value should neither be currently in use
nor reserved by the operating system. The conguration JSON le must be changed, and a new attribute
"localPort": port_number added to the interface to be modied (see example on Listing 15.2). If the
port cannot be used for any reason, the device will not be detected, resulting in a ‘No device detected’
error.
1 {
2 "network ":
3 {
4 "version ": 1,
5 "interfaces ": [
6 {
7 "address ": "172.17.1.7" ,
8 "port": 3509 ,
9 "localPort": 13485
10 }
11 ]
12 }
13 }
Listing 15.2: JSON le trying to force the local UDP port.
15.3 Connection to multiple fusionTrack devices
The new ftkInitExt function also allows to use the same library handle to connect to more than one
fusionTrack device. The recommended way to do it is to use one dedicated Ethernet adapter per fusionTrack
device, in which case, all devices must have a different IP address, if possible each on a different subnet.
Then the host PC must have one direct connection of each fusionTrack. The conguration le to set one
device with IP 172.17.1.7 and another one with IP 172.17.10.7 is presented on Listing 15.3. It is also
possible to use a GigaBit switch, implying all fusionTrack device having a different IP address in the same
subnet.
1 {
2 "network ":
3 {
4 "version ": 1,
5 "interfaces ": [
6 {
7 "address ": "172.17.1.7" ,
8 "port": 3509
9 },
10 {
11 "address ": "172.17.10.7" ,
Atracsys / 2020-06-17 / 16:48:00 85 / 113