EasyManua.ls Logo

Audio Precision B Series - Page 42

Audio Precision B Series
64 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Characteristic Specifications Supplemental Information
26 APx511 B Series Hearing Instrument Analyzer: Specifications
THD+N Measurement (Microphone, Speaker Voltage, Telecoil Current)
Fundamental Range 100 Hz to 20 kHz
Measurement Range 0 to 100%
Accuracy ±0.5 dB
Residual THD+N
1,2
(–80 dB + 1.4 µV), 20 kHz BW Microphone, speaker and telecoil
Level & THD+N Filters
High-Pass 5 Hz to 500 Hz, or None 1 Hz steps
Low-Pass
4
2.7 kHz to 90 kHz, or None 100 Hz steps; very sharp roll-off charac-
teristic exceeds AES-17
Weighting A-wt, C-wt, CCIR-1k, CCIR-2k,
CCITT, C-message, 50 μs or 75 μs
de-emph (with and without A-wt), or
None
Weighting filter is cascaded with the
high-pass and low-pass bandwidth limit-
ing filters
IMD Measurement
Test Signal Compatibility
SMPTE & MOD Any combination of 40 Hz–1 kHz (LF)
and 240 Hz–21.6 kHz (HF), mixed in
any ratio from 1:1 to 10:1 (LF:HF)
LF tone must be 1/6 • HF tone.
DFD & CCIF Any two-tone combination with mean
frequency of 250 kHz–21.56 kHz and
a difference frequency of 80 Hz–
2.0 kHz
F
mean
= (F1 + F2)/2.
F
diff
= |F2 - F1|
F
mean
must be 6 • F
diff
IMD Measured
SMPTE Amplitude modulation of HF tone Measurement BW is 40–750 Hz
MOD d2, d3, d2+d3, or d2+d3+d4+d5 Use “d2+d3” for measurements per
IEC60268
DFD d2, d3, d2+d3, or d2+d3+d4+d5 Use “d2+d3” for measurements per
IEC60268
CCIF d2 only “CCIF” is an archaic form of DFD that
measures only the d2 product. CCIF
uses a different 0 dB reference giving
readings 2x higher than DFD.
Measurement Range 0 to 20%
Accuracy ±0.5 dB

Other manuals for Audio Precision B Series

Related product manuals