CALIBRATION
AUTO-OFF AND BACKLIGHT SETTING FUNCTION
1. Prepare standard weights, and make the scale with no load.
2. When the scale is off, press and hold down TARE/UNIT button, and then press ON/OFF/ZERO
button at the same time until
appears. Release the two buttons.
3. Be sure no weight on the scale, and press ON/OFF/ZERO button, 0 in CAL-0 will be flashed.
4. When
is displayed, press and hold TARE/UNIT button to choose the unit (kg or lb)
of standard weight in calibration; load full capacity weight on the platter. Then press
ON/OFF/ZERO button, F in
will be flashed.
5. When
appears again, remove the weight from the platter, and then press ON/OFF/
ZERO button. “0” in
will flash.
6. When calibration has been completed properly, the scale will reset to normal working mode.
If
displays, that means there is an error in calibration, and needs to be re-calibrated.
7. In step 3-5, press and hold ON/OFF/ZERO to exit during the calibration mode.
1. When the scale is on, press and hold TARE/UNIT and ON/OFF/ZERO buttons till SETUP is
displayed, then the “
_XX is displayed (xx=0,1…9: means when the scale has no weight
changing and no key pressed for 1…9 minutes, the scale will auto power off to save battery
power, when x=0, means no auto-power off function). Now, press TARE/UNIT button to choose
the auto-off time and press ON/OFF/ZERO button to confirm and save the setting.
2. The scale will display “BLT.X” (x=0, 1, 2. 0 means backlight always off, 1 means backlight always
on, 2 means backlight will be on when any button is pressed or reading changes). Press TARE/
UNIT button to choose the backlight mode and press ON/OFF/ZERO button to confirm and
save the setting.