AXONICS SNM SYSTEM® Neurostimulator | Recharge-Free, Model 4101 Page 7
SPECIFICATIONS
Table 1 shows the Neurostimulator physical specifications. For detailed descriptions and specifications for other components and
accessories, refer to the product literature for those devices.
Table 1: Neurostimulator Specifications.
Physical Attributes
Height
39 mm
Length
53 mm
Thickness
6.7 mm
Weight
24.5 g
Volume
10 cc
Radiopaque identifier
RDP
Stimulation Characteristics
Frequency
2 50 Hz
Pulse Width
60 330 µs
Amplitude
0 8 mA
Minimum Amplitude Step Size
0.05 mA
Ramping
0 30 s
Stimulation Mode
Continuous or Cycling
Mode of Operation
Current-Controlled
Note:
All dimensions are approximate.
Table 2: shows the expected life of the Neurostimulator, estimated at typical stimulation settings:
Bipolar Lead Configuration, Frequency 14 Hz, Pulse Width 210 µs, continuous stimulation, impedance = 1200 .
Table 2: Neurostimulator Expected Usable Life
Stimulation Amplitude
0.5 mA
2.0 mA
3.0 mA
Life (Years)
22
9
5
Note: The neurostimulator battery may last longer than 22 years at lower energy use settings
Table 3 shows the materials used in the Neurostimulator kit components that come in contact with human tissue.
Table 3: Human-Contact Materials.
Device
Component
Material
Neurostimulator
Neurostimulator case
Titanium
Neurostimulator header
Epoxy
Septum and strain relief
Silicone
Setscrew
Titanium
Adhesive
Silicone
Torque wrench
Torque wrench handle
Polyetherimide
Torque wrench shaft
Stainless steel
Note:
The Neurostimulator case, which contains the electronics and power
source, is hermetically sealed.