Instruction for use
19
En
PRO-36
Table 2
For all ME EQUIPMENT and ME SYSTEMS
Guidance and manufacturer’s declaration – еlectromagnetic immunity
The PRO-36 is intended for use in the electromagnetic environment specifi ed below.
The customer or the user of the PRO-36 should assure that it is used in such an environment.
Immunity test IEC 60601 test level Compliance level Electromagnetic enviroment-guidance
Electrostatic discharge
(ESD)
IEC 61000-4-2
± 6 kV contact
± 8 kV air
± 6 kV contact
± 8 kV air
Floors should be wood, concrete or ceramic tile.
If fl oors are covered with synthetic material, the
relative humidity should be at least 30%.
Electrical fast transient/
burst IEC 61000-4-4
± 2 kV for power supply
lines
± 1 kV for input/output
lines
± 2 kV for power supply
lines
Mains power quality should be that of a typical
commercial or hospital environment.
Surge
IEC 61000-4-5
± 1 kV line(s) to line(s)
± 2 kV line(s) to earth
± 1 kV line(s) to line(s)
± 2 kV line(s) to earth
Mains power quality should be that of a typical
commercial or hospital environment.
Voltage dips,
short
interruptions and
voltage variations
on power supply
input lines
IEC 61000-4-11
<5 % U
T
(>95 % dip in U
T)
for 0.5 cycle
40 % U
T
(60 % dip in UT)
for 5 cycles
70 % U
T
(30 % dip in UT)
for 25 cycles
<5 % U
T
(>95 % dip in UT)
for 5 s
<5 % U
T
(>95 % dip in UT)
for 0.5 cycle
40 % U
T
(60 % dip in UT)
for 5 cycles
70 % U
T
(30 % dip in UT)
for 25 cycles
<5 % U
T
(>95 % dip in UT)
for 5 s
Mains power quality should be that of a typical
commercial or hospital environment. If the user of
the PRO-36 requires continued operation during
power mains interruptions, it is recommended that
the PRO-36 be powered from an uninterruptible
power supply or a battery.