Instruction for use
20
En
PRO-36
Power frequency
(50/60 Hz)
magnetic fi eld
IEC 61000-4-8
3 A/m 3 A/m
Power frequency magnetic fi elds should be at
levels characteristic of a typical location in a typical
commercial or hospital enviromet.
Table 3
For ME EQUIPMENT and ME SYSTEMS that are not LIFE-SUPPORTING
Guidance and manufacturer’s declaration – electromagnetic immunity
The PRO-36 is intended for use in the electromagnetic environment specifi ed below. The customer or the user of the PRO-36 should
assure that it is used in such an environment.
Immunity test
IEC 60601
test level
Compliance
level
Electromagnetic enviroment-guidance
Conducted
RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
3 Vrms
150 kHz to
80 MHz
3 V/m
80 MHz
to
2,5 GHz
3 Vrms
3 V/m
Portable and mobile RF communications equipment should be used no closer to any part
of the PRO-36, including cables, than the recommended separation distance calculated
from the equation applicable to the frequency of the transmitter
Recommended separation distance
d=1.2 P
d=1.2 P 80 MHz to 800 MHz
d=2.3 P 800 MHz to 2,5 GHz
where P is the maximum output power rating of the ransmitter in watts (W) according
to the transmitter manufacturer and d is the recommended separation distance in
metres (m).
Field strengths from fi xed RF transmitters, as determined by an electromagnetic site
survey
a
, should be less than the compliance level in each frequency range
b
.
Interference may occur in the vicinity of equipment marked with the following
symbol: