Page #9 background image
Technical Specications
EdgeGARD
®
VF
MODEL NUMBER EG4VF EG5VF EG6VF
Size (nominal) 4’ 5' 6'
Exterior Dimensions w/ Stationary
Stand (w x d x h [min to max])
53
1
2
x 32
x
[
89
1
16
- 97
9
16
”]
65
1
2
x 32
x
[
89
1
16
- 97
9
16
”]
77
1
2
x 32
x
[
89
1
16
- 97
9
16
”]
Interior Dimensions (w x d x h) 46” x 25
1
8
x 31
13
16
58” x 25
1
8
x 31
13
16
70” x 25
1
8
x 31
13
16
Shipping Weight
Benchtop model
With channel stand
435 lbs.
507 lbs.
499 lbs.
576 lbs.
564 lbs.
646 lbs.
Access Opening (working position) 14” 14" 14"
Access Opening
(loading/cleaning position)
30
3
8
30
3
8
30
3
8
Electrical Requirements 115 V, AC 115 V, AC 115 V, AC
Amps / Breaker 16 / 20 16 / 20 16 / 20
PAGE 7
General Specifications
Interior workspace made from 16-gauge stainless steel.
ISO Class 4 (Class 10)* air cleanliness in the workspace.
16-gauge cold-rolled steel exterior construction with white powder coated finish.
HEPA filters with 99.99% minimum efficiency in capturing 0.3 micrometer particles.
Separate switches for blower and light; if blower is not on, light switch will not work.
Large, easy-to-read magnehelic gauge confirms cabinet operation.
Washable pre-filter extends HEPA filter life.
One duplex on left-hand side (option for one on right-hand side).