– 15 –
Risks of single masks:
Single masks are still approved for nitrous oxide treat-
ments, but the following risks should be considered:

oxide occupational exposure limit values (refer to
Section 6. Safety and regular checks, (e.g. 100 ppm in
Germany, refer to TRGS 900) are dicult with single
masks. That is why Baldus® does not oer single masks.

-
wendung in der Zahnheilkunde [Use of Nitrous Oxide
in Dentistry], 2014).
Delivery of gas cylinders
Gas cylinders can be obtained from Baldus Medizintech-
nik GmbH.
To place a gas order:
Tel.: 0049 261/ 96 38 926-0
Fax: 0049 261/ 96 38 926-21
Email: gas@baldus-medizintechnik.de
Use tested compressed gas cylinders only! For a decen-
tralised nitrous oxide system, order at least two 10-litre
oxygen cylinders and one 10-litre nitrous oxide cylinder
for safety reasons. The nitrous oxide cylinder must be
stored upright.
Always keep a spare oxygen cylinder on hand!
For a centralised gas supply, the gas cylinders and equip-
ment are located in a separate room and the gas ows to
the treatment rooms through medical copper pipes. Bal-
dus Medizintechnik GmbH, which is ISO 13485 certied,
can also install a centralised gas supply; 50-litre cylinders
are usually used.
The cylinders must stand securely and be protected from
tipping over using straps, for example (e.g. on a cylinder
trolley with straps, see photo).
Connecting pressure reducers
to the gas cylinders
It is not possible to mix up the nitrous oxide and oxy-
gen pressure reducers to the cylinders when connecting
them, as the connectors have dierent kinds of threads.

grease and NEVER use tools to tighten them. Tightening
by hand is sucient.
Do not open the cylinders too quickly, as otherwise the
pressure exerted on the pressure reducers and O-rings
may be too strong and these may have to be replaced
sooner. Spray the pressure ports with a leak detection
spray (solution of soap and water, see Accessories) to
identify any leaks.

INSTRUCTIONS
Fig. 11: Cylinder trolley
Fig. 12: Pressure reducer (left: N
2
O, right: O
2
)

Questions and Answers: