BANTE 900P - DO mg;L or ppm Calibration; Two-Point DO (mg;L;ppm) Calibration

BANTE 900P
56 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Bante900P Portable Multiparameter Water Quality Meter
35
1.5 Dip the dissolved oxygen probe into the air-saturated water for 3 to 5 minutes.
1.6 Press ENTER key to confirm, “Calibration” indicator begins flashing.
1.7 Wait for the measured value to stabilize, the display will show “END”. The meter returns to measurement mode automatically. Single point calibration
is completed.
2 POINTS CALIBRATION:
2.1 Make sure that you have selected 2 points calibration in the setup menu.
2.2 Press CAL key, the meter shows “8.25mg/L/CAL1” (@25
o
C).
2.3 Press or key until the display shows “0.00mg/L/CAL1”.
Enter