EasyManuals Logo

Barco MDCC-4430 User Manual

Barco MDCC-4430
60 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #49 background imageLoading...
Page #49 background image
R5911227 /02 Coronis Fusion 4MP / 6MP 49
If this equipment does cause harmful interference to, or suffer from harmful interference of, surrounding
equipment, the user is encouraged to try to correct the interference by one or more of the following measures:
Reorient or relocate the receiving antenna or equipment.
Increase the separation between the equipment and receiver.
Connect the equipment into an outlet on a circuit different from that to which the receiver is connected.
Consult the dealer or an experienced technician for help.
Electromagnetic immunity
The Coronis Fusion 4MP / 6MP is intended for use in the electromagnetic environment specified below. The
customer or the user of the Coronis Fusion 4MP / 6MP should assure that it is used in such an environment.
Immunity test
IEC 60601 test levels
Compliance level
Electromagnetic
environment guidance
Electrostatic discharge
(ESD)
IEC 61000-4-2
± 8 kV contact
± 2 kV, ± 4 kV, ± 8 kV, ± 15
kV air
± 8 kV contact
± 2 kV, ± 4 kV, ± 8 kV, ± 15
kV air
Floors should be wood,
concrete or ceramic tile. If
floors are covered with
synthetic material, the
relative humidity should be
at least 30%
Electrical fast transient/
burst
IEC 61000-4-4
± 2 kV for power supply
lines
± 1 kV for input/ output
lines
100 kHz repetition
frequency
± 2 kV for power supply
lines
± 1 kV for input/ output
lines
100 kHz repetition
frequency
Mains power quality
should be that of a typical
commercial or hospital
environment
Surge
IEC61000-4-5
Line to line: ± 0.5 kV, ± 1
kV
Line to ground: ± 0.5 kV, ±
1 kV, ± 2 kV
Line to line: ± 0.5 kV, ± 1
kV
Line to ground: ± 0.5 kV, ±
1 kV, ± 2 kV
Mains power quality
should be that of a typical
commercial or hospital
environment
Voltage dips, short
interruptions and voltage
variations on power supply
input lines
IEC 61000-4-11
0% residual voltage for 0.5
period at 0°, 45°, 90°,
135°, 180°, 225°, 270°
and 315°
0% residual voltage for 1
period at
70% residual voltage for
25 periods at
Voltage interruptions: 0%
residual voltage for 250
periods at
0% residual voltage for 0.5
period at 0°, 45°, 90°,
135°, 180°, 225°, 270°
and 315°
0% residual voltage for 1
period at
70% residual voltage for
25 periods at
Voltage interruptions: 0%
residual voltage for 250
periods at
Mains power quality
should be that of a typical
commercial or hospital
environment. If the user of
the Coronis Fusion 4MP /
6MP requires continued
operation during power
mains interruptions, it is
recommended that the
Coronis Fusion 4MP /
6MP be powered from an
uninterruptible power
supply or a battery
Power frequency (50/60
Hz) magnetic field
IEC 61000-4-8
30 A/m Not applicable
1
Power frequency
magnetic fields should be
at levels characteristic of a
typical location in a typical
commercial or hospital
environment
Conducted RF
IEC 61000-4-6
3 Vrms (6 Vrms in ISM
bands)
150 kHz to 80 MHz
3 Vrms (6 Vrms in ISM
bands)
-
Radiated RF
IEC 61000-4-3
3 V/m
80 MHz to 2.7 GHz
3 V/m
Important information
1: Coronis Fusion 4MP / 6MP doesnt contain susceptible components to magnetic fields

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Barco MDCC-4430 and is the answer not in the manual?

Barco MDCC-4430 Specifications

General IconGeneral
BrandBarco
ModelMDCC-4430
CategoryMonitor
LanguageEnglish

Related product manuals