EasyManua.ls Logo

bbluv Ora - Guidance and Manufacturers Declaration - Electromagnetic Immunity; Compliance Level

bbluv Ora
72 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
17
GUIDANCE AND MANUFACTURER’S DECLARATION - ELECTROMAGNETIC IMMUNITY
IMUNITY TEST IEC 60601-1-1-2 TEST LEVEL COMPLIANCE LEVEL
Electrostatic
discharge (ESD)
IEC 61000-4-2
±8 kV contact
±2 kV, ±4kV, ±8 kV,
±15 kV air
±8 kV contact
±2 kV, ±4kV, ±8 kV,
±15 kV air
Electrical fast
transient/burst IEC
61000-4-4
Not applicable Not applicable
Surge IEC
61000-4-5
Not applicable Not applicable
Voltage dips,
short interruptions
and voltage
variations on
power supply
input lines IEC
61000-4-1-1
Not applicable Not applicable
Power frequency
magnetic eld
IEC 61000-4-8
30 A/m
50Hz/60Hz
30 A/m
50Hz/60Hz
Conducted RF
IEC 61000+4+6
Not applicable Not applicable
Radiated RF IEC
61000-4-3
10 V/m
80 MHz - 2,7 GHz
80 % AM at 1 kHz
10 V/m
80 MHz - 2,7 GHz
80 % AM at 1 kHz