BD FACSCanto II Flow Cytometer
Quick Reference Guide
This guide contains instructions for both BD FACSCanto™ clinical and
BD FACSDiva™ software. Follow only those sections that apply to the
software you are running.
Starting Up
1 Turn on the cytometer main power.
2 Start up the computer, launch the software, and
log in.
3 Make sure the software is connected to the cytometer; if needed,
connect manually (choose Cytometer or Instrument > Connect).
4 Check fluid levels in the Status or Instrument window.
Empty the waste or replace fluid containers, if needed.
5 If automatic cleaning is disabled, run the fluidics startup procedure.
6 Check the flow cell for air bubbles; if needed, run the de-gas flow cell
procedure.
7 Check that laser warmup has finished.
To prevent fluid overflow, make sure there is no tube on the sample
injection tube (SIT) during fluidics startup.
or
BD FACSCanto clinical software
BD FACSDiva software
BD FACSDiva software
BD FACSCanto clinical software