Programming Auto-Off Time
Your scale has been preset at the factory to shut down after one minute of no use.
This can be programmed from 12 seconds to 25 minutes to suit your needs.
1. Depress ON/ZERO and wait for the display to show “0.0”. Do not stand on
platform.
2. Hold MENU button down and depress the RECALL button. “Set-0” will be
displayed along with the up and down arrows; the “0” will be flashing.
3. Depress the button repeatedly until “Set-3” appears, depress ENTER.
4. The Auto-Off time currently programmed in the scale will display as minutes and
tenths of minutes (i.e. “03.2”); the far left digit will be flashing. Use the and
buttons to set the desired value, press ENTER to advance to the next digit. For
example, setting the time to “00.5” will turn the scale off after 30 seconds of no
use; “02.0” will turn the scale off after 2 minutes; “25.0” will turn the scale off
after 25 minutes.
5. Press ENTER when complete to save and return to normal weighing operation.
NOTE: If an AC adapter is used to power the scale, the scale will remain on at all
times and the Auto-Off function is disabled.
Battery Life Indicators
Figure 9: Battery indicators
Programming Time & Date
1. Depress ON/ZERO and wait for the display to show “0.0”. Do not stand on
platform.
2. Hold MENU button down and depress the RECALL button. “Set-0” will be
displayed along with the up and down arrows; the “0” will be flashing.
3. Press once, “Set-1” appears, depress ENTER.
4. Use the , and ENTER buttons to set the time, date and year.
5. Press ENTER when complete to save and return to normal weighing operation.
The scale has excellent battery life with most users getting more than two years
from a single set of batteries. Your battery longevity will vary with the frequency
of use and the length of the Auto-Off timer. To help ensure you are never
without the use of your scale, it comes with a battery life indicator. When the
battery pictorial is empty and the frame is flashing, replace batteries within a
couple weeks (Fig. 9).
NOTE: Properly dispose of old batteries. Install four (4) new alkaline D-cell
batteries. Do not mix new and used batteries.