Operation
Maintenance
1. Ascertain a proper gas supply pressure of 50psi.
2. Inspect machine hoses and connections for damage, wear and
leaks daily.
3. Perform functional test periodically.
4. Keep unit clean. Unit and accessories may be cleaned with activated
dialdehyde (Cidex). Follow manufacturers directions for use.
Service: All Service and repair must be accomplished at Belmed, Inc.
Have your dealer return unit to our facility for service.
1. Turn unit on (100% O2) and set the oxygen ow rate to equal total gas
ow (minute volume) to be administered to patient. Maintain breathing
bag about ⅔ full.
2. Apply nasal hood to patient.
3. Gradually introduce nitrous oxide ow rate while proportionally
decreasing oxygen ow rate (maintaining total gas ow) until determined
patient ratio is reached.
4. If oxygen is required, press oxygen ush button.
5. To remove patient from conscious sedation, return to 100% oxygen ow
rate established at beginning of procedure.
5