10.3 Limitations of Total Cholesterol Test
BeneCheck PLUS Total Cholesterol Test Strips give accurate results when the
following limitations are observed :
Hematocrit (Hct) percentage may affect test result. If Hct levels less than 35%
may cause incorrect high measurement results and Hct levels greater than 50%
may cause incorrect low measurement results. Please consult your medical
professionals if you do not know your Hct level.
Use only fresh capillary nger whole blood. Do not use serum or plasma.
BeneCheck PLUS Total Cholesterol Test Strips may be used at altitude up to 8,000
feet without an effect on test results.
Additional Information for Healthcare Professionals :
Interferences : Acetaminophen, uric acid, ascorbic acid, dopamine, gentistic acid,
ibuprofen, methyl DOPA, salicylate (when at physiological or therapeutical levels)
do not significantly affect the results. Hemoglobin at physiological levels ≤10 mg/dL.