Page 23Page 22
PRECELLYS EVOLUTION
PRECELLYS
®
Evolution Touch
USER MANUAL
3.7 Standardisation requirements
This equipment complies with the requirements of the CE marking except
for requirement 6.3.2 b) of IEC 61010-1. However, with respect to this
requirement, the equipment complies with the Low Voltage Directive and
the standard governing IT equipment EN 62368-1.
3.7.1 Intensive operation
If the device is used intensively (high speed, long cycle times, full tube
load, sequenced cycles), the normal rise in the overall temperature of
the system may trigger the thermal safety device. In this case, the power
supply to the motor is automatically cut o in order to prevent the device
from overheating.
3.8 Device reference
Reference of Precellys® Evolution Touch: 23405-300-RD001.
4. TRANSPORT AND STORAGE
4.1 Transport
Avoid violent shocks which could make the device not work properly.
Do the following each time you transport the device:
1- Reposition the tube holder locking foam.
2- Close the lid.
3- Use the items from the original packaging.
4.2 Storage
The device must be stored in a dry place and at a temperature between +10°C
and +40°C.
5 INSTALLATION
5.1 Unpacking the device
1. Check the contents of the box using the following list:
Should the delivery be incomplete, contact your local dealer immediately.
* A suitable adapter or cable (not supplied) is required in countries using power
outlets other than: Europe, United Kingdom, United States, Switzerland and
Australia.
1 user manual (French OR English)
1 PRECELLYS® Evolution Touch (including 1 blocking plate, 1
vacuum joint and 1 containment seal)
1 mains cable*
1 replacement vacuum joint
1 replacement mains fuse
1 individual quality control form
1 CE certicate
1 manufacturer’s inspection certicate
Do not connect the device to the power supply until
you have completed the installation.
Do not tilt the device: the Precellys® Evolution Touch
must always be resting on its four feet to avoid
damaging internal components or breaking the plastic
casing.