9 MAINTENANCE AND SERVICING
9.1 Repair
The equipment must be cleaned and decontaminated in accordance with
the procedure described in § 8.2 before shipment for repair.
The equipment must be sent with a return form indicating clearly, at the
very least, the procedure used, the persons involved and the date on which
cleaning and/or decontamination was performed. (see appendix 1).
If the return form is missing from the consignment, Bertin Technologies
after-sales team reserves the right to refuse to perform repairs.
9.2 Problems encountered
8 CLEANING AND DESINFECTION
8.1 Recommendations
For safety reasons, and to avoid damaging the equipment, the following
recommendations must be observed:
The exterior of the equipment can be cleaned with a sponge or cloth that
has been moistened with water or 70% alcohol.
8.2 Disinfection protocol
Should a tube burst during operation, for example, the parts likely to be
contaminated by infectious agents must be disinfected with a suitable
disinfectant.
In case of possible contamination of the instrument, the external parts
of the instrument (shell, cover, screen, grinding housing) can be cleaned
with a cloth moistened with water and then decontaminated with a
cloth moistened with a bleach solution (1000ppm/6°Cl), or an anionic
disinfectant. The use of ethanol can be considered but should be avoided
on rubber seals.
The user shall assume full responsibility for their choice of disinfection
process.
If another disinfection process must be implemented, contact the
technical support service beforehand in order to conrm that the
procedure is compatible with the equipment.
Do not spray liquids directly onto the device, especially the
electrical connectors and the openings of the casing (vents and
air inlet).
Unplug the equipment before cleaning.
Do not use a scraper sponge: risk of damage to the device.
Do not use caustic soda or acetone: the device will be
irreversibly damaged.
Immediately remove any liquid from the equipment with a dry
cloth.
Encountered issues Possible cause(s) Remedial measures
There is no power at the mains
socket.
1.Check the mains voltage.
2.Check that the voltage supplied by
the mains matches that of the device.
3.Check the connection of the device
to the mains.
The fuse is defective. Replace the fuse.
The ventilation system is defective.
1.Switch o the device .
2.Contact the technical support..
There is no power at the mains
socket.
1.Check the mains voltage.
2.Check that the voltage supplied
by the mains matches that of the
device.
3.Check the connection of the
device to the mains..
The fuse is defective. Replace the fuse..
The display system is defective.
1.Switch o the device.
2.Contact the technical support.
The cap was screwed on incorrectly
or the tube is defective.
The tube used is not a Precellys® tube
The operating limits were not
respected
For safety reasons, the blocking
plate stays locked in position by the
vacuum system.
Switch the device back on in order to
remove the blocking plate.
The fan is not turning.
The screen does not turn on.
A tube is no longer sealed.
If the product contained in the tube
is hazardous or potentially hazardous,
perform the appropriate disinfection
procedure.
Power cut

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Bertin Technologies PRECELLYS Evolution Touch and is the answer not in the manual?

Bertin Technologies PRECELLYS Evolution Touch Specifications

General IconGeneral
BrandBertin Technologies
ModelPRECELLYS Evolution Touch
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals