9.3.1 vacuum joint replacement
The vacuum joint located on the metal holder wears
out during operation. It is recommended to change it
when the «Pressure error» alarm appears permanently or
recurrently. As a preventive measure, this seal should be
replaced at least once every 6 months.
9.3.2 Containment seal replacement
The containment seal should be changed at least once a year. The tube
holder does not have to be dismantled when replacing it.
Switch o the appliance and unplug
the power cord before carrying out
this operation.
1. Place the inner groove of the containment seal over the
metal ange.
2. Place the outer groove of the containment seal on
the housing. The containment seal should be changed
at least once a year. This procedure does not require the
dismantling of the metallic support ring.
9.3.3 Fuse replacement
Switch o the appliance and unplug the power cord
before carrying out this operation.
Remove the fuse holder by pinching the retaining clips .
9.3 Wear parts replacement
This paragraph species the maintenance that users must perform on a
regular basis to ensure that the Precellys® Evolution Touch works properly
and that the tubes are maintained during the homogenization phase:
The voltage levels in the device are hazardous. In order
to guarantee the safety of the user, including during the
cleaning and disinfection phases, the containment seal
must not be damaged (puncture, tear, etc.)
DO NOT tilt the device: the Precellys® Evolution Touch must always be
resting on its four feet. If not, the internal components can get damaged
or the plastic casing will break.
Wearing part Reference
Frequency of
replacement
Why ?
2 mL blocking plate +
tube holder
P002274-PEVT0-A.0 when damaged
Necessary for maintaining
tubes in place during
homogenization.
7 mL blocking plate +
tube holder
P000911-PEVT0-A.0 when damaged
Necessary for maintaining
tubes in place during
homogenization.
15 mL blocking plate
+ tube holder
P000810-PEVT0-A.0 when damaged
Necessary for maintaining
tubes in place during
homogenization.
Vacuum joint P000806-PEVT0-A.0
6 months or when
damaged
Necessary for maintaining
tubes in place during
homogenization.
Containment seal P000805-PEVT0-A.0
1 year or when
damaged
Necessary to guarantee the
tightness of the device.
Mains fuse P002574-PEVT0-A.0 when damaged
Anti-rotation kit P002378-PEVT0-A.0
1 year or when
damaged
Necessary for maintaining
tubes in place during
homogenization.
at screwdriver

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Bertin Technologies PRECELLYS Evolution Touch and is the answer not in the manual?

Bertin Technologies PRECELLYS Evolution Touch Specifications

General IconGeneral
BrandBertin Technologies
ModelPRECELLYS Evolution Touch
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals