Page 13Page 12
PRECELLYS EVOLUTION
PRECELLYS
®
Evolution Touch
USER MANUAL
1
2
3
Protocol parameters Values
Speed*
From 4500 to 6800 rpm and from 7200 rpm
to 10 000 rpm increments of 100 rpm
Number of cycles From 1 to 10
Grinding time 5 to 120 s by increment of 1 second
Time delay between 2 periods 5 to 120 s by increment of 1 second
Tube size From 0.3 mL to 50 mL
** Cooling module ON or OFF
** Setpoint temperature From 0 to 10°C by increment of 1°C
** Cooling control mode Manual or automatic
3.1.4.1 Programming
The PRECELLYS® Evolution Touch is designed to operate at a maximum
speed of 10 000 rpm.
A PRECELLYS® Evolution Touch operating protocol is made up of several
successive cycles. Each protocol is composed of dierent parameters
listed in the table here-after and their setting can be made in the “Protocol
settings” or “Protocol library” menus.
* Programming speed between 6900 rpm and 7100 rpm is not permitted,
in order to ensure good functioning and a maximum product life.
** The cooling module is an option.
The operating protocol is under the responsibility of the operator. The
manufacturer recommends for each speed an operating range that
prevents a strong rise in the temperature of the device, the following
paragraph gives the limits of the adjustment parameters accessible to the
operator.
LED: indicates the status of the equipment:
• green: equipment operational and ready to start a cycle
• red: alarms
Screen: displays all the information related to the settings and
faults of the equipment
Start/Stop Button: starts and stops the cycle
1
2
3
Access to the
protocol settings.
Access to the list of
saved protocols
Access to the settings
of the device.
3.1.4.2 Description of the command interface