TECHNICAL SPECIFICATIONS
Insulation: Class2, Double insulation.
Overvoltage category: CATIV 600V
Note:
These meters meet CAT III and CAT IV IEC 61010 standards. The IEC 61010 safety standard
denes four overvoltage categories (CAT I to IV) based on the magnitude of danger from transient
impulses. CAT III meters are designed to protect against transients in xed-equipment installations
at the distribution level; CAT IV meters are designed to protect against transients from the primary
supply level (overhead or underground utility service).
Maximun voltage between any terminal and earth ground: 600V DC/AC RMS
Display: 11000 counts LCD display
Polarity: Automatic, (-) negative polarity indication
Over-range: “OL” mark indication.
Low battery A battery symbol is displayed when the battery voltage drops
indication: below the operating level.
Measurement rate: 4 times per second nominal.
Auto power off: Meter automatically shuts down after approx. 15 minutes of inactivity.
Operating environment: -10 °C to 50 °C (14 °F to 122 °F) at < 70 % relative humidity.
Storage temperature: -30 °C to 60 °C (-4 °F to 140 °F) at < 80 % relative humidity.to
Relative humidity: 90% (0°C to 30°C); 75%(30°C to 40°C); 45%(40°C to 50°C);