46
EN
All AC voltage ranges are specied from 5% of range to 100% of range
Input Impedance: 9.7MΩ.
AC Response: 50 Hz 60Hz
Maximum Input: 600V dc or 600V ac rms.
DC Current (auto-ranging)
Range Resolution Accuracy
110.00µA 0.01µA
±1,5% of rdg ±5 digits
1100.0µA 0.1µA
11.000mA 1µA
±1,5% of rdg ±10 digits
110,00mA 10µA
10.000A 1mA ±2,5% of rdg ±10 digits
Overload Protection: 0.2A / 600V and 10A / 1000V Fuse.
Maximum Input: 110µA dc on µA range
110mA dc on mA range
10A dc on 10A range.