34
Meas-
urement
method
Oscillometric, non-invasive blood pressure
measurement on the upper arm
Measure-
ment range
Cu pressure 0 – 300 mmHg,
systolic pressure 50 – 280 mmHg, diastolic
pressure 30 – 200 mmHg,
pulse 40 – 199 beats/minute
Display
accuracy
Systolic pressure ± 3 mmHg,
diastolic pressure ± 3 mmHg,
pulse ± 5% of the displayed value
Meas-
urement
uncertainty
Max. permissible standard deviation accord-
ing to clinical testing:
systolic pressure 8 mmHg,
diastolic pressure 8 mmHg
Memory 4 x 30 memory spaces
Dimensions L 112 mm x W 110 mm x H 53 mm
Weight Approx. 341 g (without batteries, with cu)
Cu size 22 to 42 cm upper arm circumference
Operating
conditions
+ 10 °C to + 40 °C, <90% relative humidity,
800 –1050 hPa ambient pressure
Storage and
transport
conditions
-20 °C to + 55 °C, <90% relative humidity
(non-condensing)
Power
supply
4 x 1.5 V
LR6 AA batteries
Battery life For approx. 300 measurements, depending
on blood pressure and inflation pressure
levels
Expected
service life
Information on the service life of the product
can be found on the homepage
Classifica-
tion
Internal power supply, IP 21, no AP or APG,
continuous operation, type BF applied part
The serial number is located on the device or in the battery com-
partment.
We reserve the right to make technical changes to improve and
develop the product.
• This device conforms with the European standard
EN 60601-1-2 (Group 1, Class B, in accordance with
CISPR-11, IEC 61000-3-2, IEC 61000-3-3, IEC 61000-4-2,
IEC 61000-4-3, IEC 61000-4-4, IEC 61000-4-5,
IEC 61000-4-6, IEC 61000-4-7, IEC 61000-4-8,
IEC 61000-4-11) and is subject to particular precautions with
regard to electromagnetic compatibility. Please note that