42
Cu size 22 to 42 cm upper arm circumference
Operating
conditions
+ 10 °C to + 40 °C, < 90 % relative humidity,
800 – 1050 hPa ambient pressure
Storage
conditions
-20 °C to + 55 °C, < 90% relative humidity
(non-condensing)
Power
supply
4 x 1.5 V
AA batteries
Battery life Lasts for approx. 300 measurements de-
pending on the blood pressure and inflation
pressure as well as the number of Bluetooth
®
connections
Expected
service life
At least 20.000 measurements
Classifica-
tion
Internal power supply, IP20 no AP or APG,
continuous operation, type BF applied part
Data
transfer via
Bluetooth
®
wireless
technology
The device uses Bluetooth
®
,
Frequency band 2.400 – 2.483 GHz, max.
transmission power 5 dBm,
The serial number is located on the device or in the battery
compartment.
Technical specifications are subject to change without notifi-
cation to allow for updates.
• This device conforms with the European stand-
ard EN 60601-1-2 (in accordance with CISPR-11,
IEC 61000-3-2, IEC 61000-3-3, IEC 61000-4-2,
IEC 61000-4-3, IEC 61000-4-4, IEC 61000-4-5,
IEC 61000-4-6, IEC 61000-4-7, IEC 61000-4-8,
IEC 61000-4-11) and is subject to particular precautions
with regard to electromagnetic compatibility. Please
note that portable and mobile HF communication sys-
tems may interfere with this device.
• The accuracy of this blood pressure monitor has been
carefully checked and developed with regard to a long
useful life. If the device is used for commercial medical
purposes, it must be regularly tested for accuracy by ap-
propriate means. Precise instructions for checking accu-
racy may be requested from the service address.
Mains part
Model no. LXCP12X-050100BG
Input 100 – 240 V, 50 – 60 Hz, 0.5 A max
Output 5 V DC, 1A, in conjunction with Beurer
blood pressure monitors only
Manufacturer Shenzhen Iongxc power supply co., ltd.
Protection This device is double protected and has a
primary-side cutout switch which discon-
nects the device from the mains in case of
malfunction.
Ensure that you have removed the batteries
from the battery compartment before you
use the mains part.
Polarity