30
Problem Possible cause Solution
The skin
turns red
and/or a
stabbing
pain occurs.
Thecuisnotposi-
tioned correctly on
the skin.
Ensure that the
cusitssecurely
on the skin and
cannot move.
Thecuisdirty.
Cleanthecuas
described in these
instructions for
use.
One of the water
contact electrodes
has a scratch.
Replacethecu.
Water contact elec-
trodes are not su-
ciently moist.
Switchothe
device. Moisten
electrodes.
Stimulation
stops during
use.
Thecuhascome
loose from the skin.
Stop the applica-
tion and reposition
thewristcu.
The connection
cable has worked
itself loose.
Stop the applica-
tion and reinsert
the connection
cable.
The batteries are
empty.
Replace the
batteries on the
control unit.
Ifyoucannotndasolutiontoyourproblemhere,contact
ourCustomerServices.
10. Technical specifications
Wrist circumference Fromapprox.14cm
Type EM 29
Model EM 28
Weight
–Device
–Device,cuand
cable
– Approx.90gincl.batteries
–Approx.155gincl.batteries
Electrode size Approx.80cm²
Parameter
(500ohmload)
Output voltage:
max.50Vpp/5.5Vrms
Output current:
max.100mApp/11mArms
Output frequency:
2 – 110 Hz
Pulse length 60 – 220 μs per phase
Waveform Symmetric,biphasic
rectangular pulse
Voltagesupply 4.5V
(3x1.5VAAA,typeLR03)