ELECTRICAL SPECIFICATIONS/EQUIPMENT CLASSIFICATION
The Model SC-3004-DL interior components are “double insulated” and do not require a
“protecve ground.” The system is equipped with an 18 gauge, 2-wire, 10. Power cord, se-
cured through the pump casing with a Heyco strain relief brushing as well as an addional
“hold-down” clamp for added safety.
1. Class of protecon against electrical shock: CLASS II EQUIPMENT
2. The degree of protecon against electric shock: APPLIED PART-TYPE B
3. Mode: CONTINUOUS OPERATION WITH INTERMITTENT LOADING
4. According degree of protecon against ingress of water: IP20
EMC Manufacturer’s Declaraon
Model SC-3004-DL electromagnec emissions—manufacturer's declaraon
The model SC-3004-DL is intended for use in the electromagnec environment specied below. The customer or the user of the model SC-3004-DL should assure
that it is used in such an environment.
Emissions test Compliance Electromagnec environment— guidance
RF emissions
CISPR 11
Group 1 The model SC-3004-DL uses RF energy only for its internal funcons. Therefore, its RF emissions are
very low and are not likely to cause any interference in nearby electronic equipment.
RF emissions
CISPR 11
Class B
The model SC-3004-DL is suitable for use in all establishments, including domesc establishments and
those directly connected to the public low voltage power supply network that supplies building used
for domesc purposes.
Harmonic emissions
IEC 61000-3-2
Not applicable
Voltage uctuaons / icker
emissions
IEC 61000-3-3
Not applicable
Guidance and manufacturer’s declaraonelectromagnec immunity
The model SC-3004-DL is intended for use in the electromagnec environment specied below. The customer or the user of the model SC-3004-DL should assure
that it is used in such an environment.
Immunity test
IEC 60601
test level
Compliance level Electromagnec environment—guidance
Electrostac
discharge (ESD)
IEC 61000-4-2
± 2, 4, and6 kV contact
± 2, 4 and 8 kV air
± 2, 4, and6 kV contact
± 2, 4 and 8 kV air
Floors should be wood, concrete or ceramic
le. If oors are covered with synthec ma-
terial, the relave humidity should be at least
30%.
Electrostac
fast transient / burst
IEC 61000-4-4
± 2 kV for power supply lines
± 1 kV for input / output lines
± 2 kV
Not applicable
Mains power quality should be that of a
typical home use locaon.
Surge
IEC 61000-4-5
± 0.5 and 1 kV line(s) to line(s)
± 2 kV line(s) to earth
± 0.5 and 1 kV dierenal mode
Not applicable, no ground wire
Mains power quality should be that of a
typical home use locaon.
Voltage dips, short
interrupons and voltage
variaons on power supply
input lines
IEC 61000-4-11
<5 % U
T
(>95% dip in U
T
) for 0.5 cycles
40 % U
T
(60% dip in U
T
) for 6 cycles
70 % U
T
(30% dip in U
T
) for 30 cycles
<5 % U
T
(>95% dip in U
T
) for 5 s
<5 % U
T
(>95% dip in U
T
) for 0.5 cycles
40 % U
T
(60% dip in U
T
) for 6 cycles
70 % U
T
(30% dip in U
T
) for 30 cycles
<5 % U
T
(>95% dip in U
T
) for 5 s
Mains power quality should be that of a
typical home use locaon. If the user of the
model SC-3004-DL requires connued opera-
on during mains power interrupons, it is
recommended that the model SC-3004-DL be
powered from an uninterrupble power
supply or a baery.
Power frequency (50/60
Hz) magnec eld
IEC 61000-4-8
3 A/m 3 A/m The power frequency magnec elds should
be at the levels found in a typical home use
locaon.
NOTE: U
T
is the AC mains voltage prior to applicaon of the test level.