7. HOUR METER
7.1 The hour meter is displayed for 5 seconds each me you turn on your pump. The
numbers in the “LED Readout” show the 1-999 hours, the leer in the “Chamber
Number LED” indicates the following:
A. Less than 1000 hours
B.= 1000 - 2000 hours
C.= 2000 - 3000 hours
D. = 3000 - 4000 hours
E. = 4000 - 5000 hours
F.= 5000 - 6000 hours
7.2 To RESET the internal hour meter and return the pump to factory sengs, with
the pump turned o, press and hold the “Mode”, “Up” and “Down” buons all
at the same me for 5 seconds. The LED will then Light up.
7.3 Aer 5 seconds, release the three buons and press the “MODE” buon. This
will reset the internal hour meter and return the pump to the original factory
sengs.
IMPORTANT USAGE INFORMATION:
This pump has a customized soware that remembers the pressure require-
ments of each and every paent. It is very important to reset the pump back
to the original factory sengs prior the placing the pump on a new paent.
MAINTENANCE AND STORAGE:
EXTERIOR PUMP CASING CLEANING INSTRUCTIONS:
1. Clean the exterior case and tubing with a damp (not wet) cloth using mild soap
and water soluon once per month or as needed.
GARMENT CLEANING/DISINFECTING INSTRUCTIONS:
Disconnect garment from device
2. Open garment to expose all sides either by separang Velcro type hook and loop
or by unzipping (depending on type of garment).
WARNING!
Only an authorized technician may open the pump
Before cleaning, unplug power cord from electrical outlet
WARNING!
Do not allow liquids to enter the pump, as this can present an electrical
hazard
Always allow the pump to dry before using
Do not use bleach on the pump