$2C2?@2E.0A&?.0821
Reverse the Exact-Track bed. The Gait Trainer 3 bed is double-sided, allowing it to be reversed and
used over. Once both sides have been used, the bed must be replaced.
$2=9.02&?2.1/29A
Inspect Treadbelt for cracks or tears. If none are found, continue to use. If any cracks or tears are
apparent, replace treadbelt.
& "$!'$%
29A208B/?60.A6<;
(See Figure 11.4.)
1. Using the large syringe provided, squirt one-half tube of the lubricant underneath the center
of the treadbelt.
2. Walk 10 steps on the Gait Trainer 3 at a speed of 1.0 mph. This will moisten an 8” track under-
neath the center of the entire treadbelt.
3. Allow the Gait Trainer 3 to dry for approximately 10 minutes.
Use only the Biodex lubricant kit with your Gait Trainer 3. Most standard greases, waxes and sil-
icon sprays will build up on the rollers, causing belt slippage and affecting tracking.
To re-order lubricant kit, use Biodex part # 945-276. Each container provides 12 applications.
Figure 11.4. Squirt one-half tube of lubricant between the belt and deck.
CONTENTS
— 11-3 — MAINTENANCE