11. Electromagnetic Compatibility
Standard Test Method Range
Limits
Result
IEC 61000-3-2 Harmonics 100 Hz to 2KHz Class A THD = 251.94%
IEC 61000-3-3 Flicker observation time (TP) 10 min - -
max voltage change (dmax) 4% 0.00 %
max Rel steady state voltage change (dc) 3% 0.00 %
duration of d(t)>3%(t) .2 sec 0.00 sec
short term flicker Sev (PST) 1.00 0.00
long term flicker Sev (PLT) .65 0.00
IEC 61000-4-2 Electrostatic Discharge Contact: 4 & 6Kv pos/neg 1pps for 10 sec no degradation complied
Air: 2, 4, & 8Kv pos/neg 1pps for 10 sec of performance
IEC 61000-4-3 Radiated Immunity
80 MHz to 1000 MHz / 3v/m Horiz &
no degradation complied
Vertical @ 2M of performance
1000 MHz to 2500 MHz / 3v/m Horiz &
Vertical @ 1M
IEC 61000-4-4 Electrical Fast Transient/ PWR Input leads .5, 1, & 2 Kv / pos & no degradation complied
Burst, Power Leads
neg / 5KHz Rep Rate
of performance
IEC 61000-4-6 Conducted Immunity, 150 KHz to 80 MHz /3Vrms no degradation complied
Power Leads of performance
IEC 61000-4-8 Magnetic Immunity
3A/M RMS @ 50/60 Hz
no degradation complied
of performance
IEC 61000-4-11 Voltage Dips and Interrupts Int Pause % no degradation complied
Duration between reduction of performance
Int 20msec pause 10 sec
>95%
no degradation complied
of performance
Int 100msec pause 10 sec 60% no degradation complied
of performance
Int 500msec pause 10 sec 30% no degradation complied
of performance
Int 5000msec pause 10 sec >95% no degradation complied
of performance
CISPR 11 Conducted Emissions 150 KHz - 500 KHz / 5 MHz - 30 MHz, 79/73 dbuV QP complied
Class A, Group 1 66/60 dbuV AV
CISPR 11 Radiated Emissions 30 MHz - 230 MHz / 230MHz - 1GHz, 40/47 dbuVm @ complied
Class A, Group 1 10 m
IEC 61000-4-5 Surge Immunity, 1 Kv, differential 1ppm, pos/neg no degradation complied
Power Leads of performance
— 27 — ELECTROMAGNETIC COMPATIBILITY