10 11
Treatment Completion
Therapeutictreatmentshouldnotbesuspendeduntilfusionoccursoruntilsuchtime
asadeterminationismadebytheprescribingphysician.Thedeviceisprogrammedto
deliver 270 continuous days of therapeutic treatment and automatically discontinues
operationafterthe270days.
Patient Compliance Monitoring
TheBiomet
®
SpinalPak
®
Non-invasiveSpineFusionStimulatorSystemcontainsembeddedsoftware
whichallows the display of patient specific history data includingusage andtherapeutic treatment
times.Thisdatamaybedownloadedtoapersonalcomputerforviewing,storageand/orprintoutvia
theuseofBiometComplianceDataDownloadSoftware.PleasecallyourlocalBiometrepresentative
toobtainmoreinformation.
Ordering Information
Toordersupplies,contactBiomet.Seepage2“ImportantSafeguards”forcontactinformation.
The following information is necessary to expedite any inquiry:
• Patientname
• Physicianname
• Addresstosendreplacementparts(patienthome,MDoffice,etc.)
Electrode Instructions for Use
Donotopenouterpacketuntilreadytouse.
1) Tearopenpacket.
2) Removeelectrodefromclearplasticbackingliner.
3) Wetfingerwithtapwaterandmoistenentiregelarea.
4) Placeelectrodeonskin.
5) Connectelectrodetoelectrodelead.
Renewal
1) Withcontinuoususe,electrodesmaydryout.
2) Torenew,wetfingerwithtapwaterandmoistenentiregelarea.
3) Reapplyelectrodetoskin.
Storeinacoolplace