Directions for Use
RECOMMENDED USAGE
The Biomet
®
SpinalPak
®
Non-invasive Spine Fusion Stimulator System is designed to deliver 270
daysofcontinuoustherapeutictreatmentfor24hoursperday.Therecommendeddailytherapeutic
treatmentiscontinuousfor24hours.
OPERATING INSTRUCTIONS
TheBiomet
®
SpinalPak
®
Non-invasiveSpineFusionStimulatorSystemhasbeenspecificallydesigned
tobeconvenienttouse,comfortabletowear,andsafetooperate.Patientsshouldbeginusingthe
Biomet
®
SpinalPak
®
Non-invasive Spine Fusion Stimulator System immediately after reading the
instructionsforuseandhavingreceivedinstructionsfromtheirprescribingphysician.
To begin treatment - Electrode Placement
• Cleananddrytheskinwheretheelectrodeswillbeplaced.Trimming(notshaving)bodyhairfrom
theelectrodeapplicationareaisoftenhelpful.
• Note: To ensure electrode placement and adhesion, you may use one of the provided electrode
retainer patches. Place one electrode
twotothree inches totheleft of the
area of the surgery and a second
electrode two to three inches to the
rightoftheareaofthesurgerysothat
the electrodes are four to six inches
apart. See Figure 2.Dependingonthe
patient’sabilitytomoveaftersurgery,
it may be helpful for the patient to
ask another person to assist them
inplacing the electrodes. See instructions for use on Page 10. The patientshould consult their
prescribingphysicianorBiometiftheyhaveanyquestionsorconcernsregardingproperelectrode
placement.If theirskinbecomesabnormallyred attheelectrodesites,theelectrodesshouldbe
moved adjacent to the original sites. If the redness does not go away after 48 hours with the
electrodesremoved,thepatientshouldcontacttheirprescribingphysician.
Tips:
Loose electrodes–Confirmthatbothelectrodesareincompletecontactwithclean,dryskin.See
instructionsontheelectrodepacket.
Incomplete circuit/disconnection –Checkallconnectionpoints,confirmingasnugfit.
Broken electrode lead wire–Ifalarmingcontinues
afterconfirmingconnection,attachanewelectrode
leadwire.
To begin treatment - Lead Wires
• Thepatientisprovidedwithtwostimulatorleadwire
lengthswith theBiomet
®
SpinalPak
®
Non-invasive
SpineFusionStimulatorSystem.Thepatientshould
selectthelengthofthestimulatorleadwireinorder
to provide both convenience and comfort when
wearingthestimulatorduringtreatment.
• Insertthestimulatorleadwiremaleconnectioninto
eachfemaleelectrodeleadwireconnection.
• Inserttheleadwireplugintotheopeningatthetopofthestimulator.See Figure 3.
Figure 2
Figure 3